PRINTER VERSION The flow sensing grid in
the FlowHood unit is high impact ABS plastic. It can be damaged if subjected to physical
abuse or excessive stress. Accurate air flow measurements are dependent on the integrity
of this grid. Even a hairline crack will effect the results. The grid should not be
exposed to temperatures in excess of 140/°F for extended periods, or 160/°F for five
minutes. Do not lay anything on the grid or interfere with the tension of the support
springs.
Check the flow sensing grid
periodically for damage, poor connections to the meter, or an accumulation of dirt or dust
particles. The grid may be cleaned by wiping carefully with alcohol. Care must be taken to
avoid knocking dirt or dust particles into the grid orifices.
The top support assembly, consisting of the
aluminum rod structure that supports the cloth skirt and frame assembly, should be stored
folded with the ends of each pair of legs inserted into the top and middle holes of the
two front base corner tubes. The head of the assembly should be toward the back of the
base assembly. The four spring rods should be positioned downward to avoid applying
pressure to the flow sensing grid.
The fabric tops should be washed
periodically in cool water with a mild detergent. The tops must be air dried only as heat
may cause the fabric to shrink. Excessive dirt build up should be avoided.
Reasonable care will prolong the life of
the fabric tops. Sharp objects or corners can puncture the fabric and affect the accuracy
of the readings.
|