AIR FLOW:
Measured in cubic feet per
minute (cfm) and litres per second (L/s) corrected for local air
density. The measurement range is 25-2500 cfm supply and 25-1500 cfm
exhaust. Accuracy is ±3% of reading ±7 cfm from 100 to 2000 cfm.
AIR DENSITY CORRECTION:
The air density correction range is
14-40 in Hg and
32ºF to 158ºF for correction of
air flow measurements. The readings represent local density airflow
corrected for the density effects of temperature and absolute
pressure.
RESPONSE TIME:
Five seconds for air flows greater than 180 cfm, and up to eight
seconds at less than 70 cfm.
POSITION
SENSITIVITY:
Unaffected by position.
READOUT:
10 digit, 0.4", high contrast, liquid crystal display.
WEIGHT:
FlowHood unit with meter and 2'x2’ top,
9.7 lbs.
UNIT DIMENSIONS:
Height 40"; top 24"x24"; base 14"x14".
BATTERY LIFE:
3000 readings per charge, 500 recharge cycles. |