AIR VELOCITY:
±3% of reading ± 7 fpm 50 to 8,000 fpm pitot tube (30,000 fpm FS); 50 to 5,000 fpm AirFoil; 50
to 2500 fpm VelGrid.
DIFFERENTIAL PRESSURE:
±2% of reading ± 0.001 in wc from 0.0500
to 50.00 in wc, (0.0001 to 60 in wc FS); 20 psid safe pressure.
TEMPERATURE:
±0.5°F
accuracy from 32°F to 158°F.
AIR FLOW: Accuracy is ±3% of reading ±7 cfm from 100 to 2000 cfm; range is 25 to 2500 supply, 25 to 1500 exhaust with 8400 FlowHood.
ABSOLUTE PRESSURE: ±2% of reading ±0.1 in Hg from 14 to 40 in Hg
referenced to vacuum. 60 psia maximum safe pressure.
AMBIENT RANGE: Full range compensation from 40°F to 140°F.
AIR DENSITY CORRECTION: Local or standard (mass flow) air density
correction range is 14 to 40 in Hg and 32ºF to 158ºF.
POSITION SENSITIVITY:
Unaffected by position.
MEMORY: 2000 readings, sequence labeled, sum, average, minimum and maximum.
CALIBRATION: Calibration certified NIST traceable.
READOUT: 10 digit, 0.4", high contrast, liquid crystal display.
METER HOUSING: High impact, molded, "T" grade ABS.
METER WEIGHT: 38 ounces (1.08kg), including batteries
SIZE: 6.0" x 6.4" x 2.7" (15.2 x 16.3 x 6.9 cm)
CONNECTIONS: 1/4" OD slip-on for 3/16" ID soft tubing.
BATTERY LIFE: 3000 readings per charge, 500 recharge cycles. |